Exatidão e reprodutibilidade do tempo de exposição dos aparelhos de raios-x odontológicos / Accuracy and reproducibility regarding the time of x-ray dental machines exposure

Authors

  • Rafael Terças Travassos Universidade Federal do Maranhão - UFMA
  • Ana Emília Figueiredo de Oliveira Universidade Federal do Maranhão - UFMA
  • Kyria Spyro Spyrides Universidade Federal Fluminense
  • Fernanda Ferreira Lopes Universidade Federal do Maranhão - HUFMA
  • Rodrigo Ribeiro Gonçalves Pinho Universidade Federal do Maranhão - UFMA
  • Maria Aparecida Costa Universidade Federal do Maranhão - UFMA

DOI:

https://doi.org/10.18764/

Keywords:

Radiografia, Tempo de exposição, Controle de qualidade

Abstract

Introdução: O Controle de qualidade radiográfica realizada por meio da verificação periódica dos parâmetros técnicos dos equipamentos de raios-X representa um dos aspectos fundamentais para uma excelente qualidade da imagem com mínima dose de exposição ao paciente. Objetivo: Investigar parâmetros técnicos relativos ao tempo de exposição dos aparelhos de raios-X odontológicos, instalados em Faculdades de Odontologia, consultórios públicos e privados da cidade de São Luís-MA.
Metodologia: Foram registrados os dados nominais do tempo de exposição e tempos efetivos (real/apresentado no cronômetro) e o tipo de disparador do aparelho de raios-X, se mecânico ou eletrônico. Para medir o tempo de exposição do feixe de raios-X e verificar a sua exatidão, empregou-se quatro tempos de exposição nominal: 0,3, 0,5, 0,6 e 0,8 segundos. Realizaram-se quatro medições para cada tempo e dentre essas, selecionou-se o maior valor de leitura (L maior) e o menor valor de leitura (L menor) de cada tempo de exposição selecionado para realização dos cálculos do teste de reprodutibilidade do tempo de exposição. Resultados: Observou-se que apesar dos tempos terem se mostrado reprodutíveis em um grande percentual (92,5%) não repetiram o mesmo desempenho em relação à exatidão do tempo de exposição. A falta de reprodutibilidade apresentou percentual mais desfavorável no tempo de exposição mais baixo, o de 0,3s (18%), sendo minimizado sucessivamente para os tempos de 0,5s (7%), 0,6s (4%) e 0,8s (1%). Conclusão: A maioria dos aparelhos de raios-X avaliados apresentou comprometimento em relação à exatidão do tempo de exposição, com bom desempenho relativo à reprodutibilidade.

Palavras-chave: Radiografia. Tempo de exposição. Controle de qualidade

Abstract
Objective: To investigate the technical parameters relating the time of x-ray dental machines exposure, located in dentistry colleges and at public and private dental offices in the city of São Luís, MA. Methods: We recorded the nominal data of exposure time and effective time (real / showed in the timer), the type of triggering device of the x-ray, whether it was mechanical or electronic. In order to measure the exposure time of x-ray beam and check its accuracy, we performed four times of nominal exposure: 0.3, 0.5, 0.6 and 0.8 seconds. There were four measurements for each time and among these we selected the highest value of reading (higher R) and the lowest value of reading (lower R) of each selected exposure time to do calculations of the time exposure's reproducibility test. Results: Although the time periods showed to be reproducible mostly (92.5%), there was not the repeat of the same performance in relation to the accuracy of exposure time. The lack of reproducibility showed the worst
percentage in the lowest time of exposure 0.3 s (18%), being reduced successively for periods of 0.5 s (7%), 0.6 s (4%) and 0.8s (1%). Conclusion: Most of x-ray machines evaluated showed defects on the accuracy of exposure time, with good performance on the reproducibility.

keywords: Radiography. Exposure time. Quality control

Downloads

Download data is not yet available.

Author Biographies

Rafael Terças Travassos, Universidade Federal do Maranhão - UFMA

Graduando de Odontologia. Bolsista de Iniciação Científica BIC/UFMA/CNPq. Universidade Federal do Maranhão.

Ana Emília Figueiredo de Oliveira, Universidade Federal do Maranhão - UFMA

Doutorado em Radiologia Odontológica pela Universidade Estadual de Campinas. Pós-Doutorado pela Universidade da Carolina do
Norte/Chapel Hill, Estados Unidos. Professor adjunto da Universidade Federal do Maranhão - UFMA.

Kyria Spyro Spyrides, Universidade Federal Fluminense

Doutora em Radiologia Odontológica pela Universidade Estadual de Campinas. Pós-Doutorado pela University of North Carolina at Chapel
Hill, Estados Unidos. Professora da Universidade Federal Fluminense, Brasil.

Fernanda Ferreira Lopes, Universidade Federal do Maranhão - HUFMA

Doutora em Patologia Oral pela Universidade Federal do Rio Grande do Norte, Brasil. Professor Adjunto da Universidade Federal do
Maranhão - HUFMA.

Rodrigo Ribeiro Gonçalves Pinho, Universidade Federal do Maranhão - UFMA

Graduando em Odontologia. Universidade Federal do Maranhão - UFMA.

Maria Aparecida Costa, Universidade Federal do Maranhão - UFMA

Doutoranda em Odontologia. Universidade Federal do Maranhão - UFMA.

How to Cite

Travassos, R. T., Oliveira, A. E. F. de, Spyrides, K. S., Lopes, F. F., Pinho, R. R. G., & Costa, M. A. (2012). Exatidão e reprodutibilidade do tempo de exposição dos aparelhos de raios-x odontológicos / Accuracy and reproducibility regarding the time of x-ray dental machines exposure. Revista De Pesquisa Em Saúde, 12(1). https://doi.org/10.18764/

Issue

Section

Artigos Originais / Original Articles