ANÁLISE COMPARATIVA DA DETERMINAÇÃO DE GLICEMIA CAPILAR E VENOSA COM GLICOSÍMETRO VERSUS DOSAGEM LABORATORIAL / COMPARATIVE ANALYSIS OF CAPILLARY AND VENOUS GLYCEMIA WITH GLUCOMETER VERSUS LABORATORY DOSAGE

Auteurs-es

  • Sally Cristina Moutinho Monteiro Universidade Federal do Maranhão - UFMA
  • Elizabeth Gomes
  • Ilka Kassandra Belfort Universidade Federal do Maranhão - UFMA
  • Mauricio Fernandes Avelar
  • Romildo Martins Sampaio Universidade Federal do Maranhão - UFMA

DOI :

https://doi.org/10.18764/

Résumé

Introdução: Os sistemas para auto monitoramento da glicemia devem fornecer valores de glicose precisos e reprodutíveis para poder assegurar decisões terapêuticas adequadas em pessoas com diabetes. Objetivo: Realizar a avaliação comparativa entre os níveis de glicose, a partir da determinação por teste laboratorial remoto (TLR) e método laboratorial padrão. Métodos: Estudo descritivo realizado com usuários atendidos em uma Unidade Básica de Saúde. A determinação da glicemia capilar foi realizada com equipamento Accu-Check Advantage (Roche), utilizando lancetas manuais, e os níveis de glicemia venosa foram determinados com metodologia enzimática (glicose oxidadase - Labtest Diagnóstica), utilizando controle de qualidade de dois níveis. Resultados: As determinações da glicemia foram realizadas em 50 de voluntários (11 do sexo masculino e 39 do sexo feminino), com média de idade de 61,58 ± 11,08 anos, sendo que 25 participantes possuem Diabetes Mellitus do tipo 2. O coeficiente de correlação de Spearman e o teste de Bland Altman foram realizadas para comparação das metodologias. As análises estatísticas demonstraram uma ótima correlação entre níveis de glicemia capilar e venosa (r = 0.995). Conclusão: O teste laboratorial remoto é eficaz e pode contribuir com o controle dos níveis de glicemia, e também pode minimizar agravos à saúde e melhoria da qualidade de vida dos diabéticos.

Palavras-chave: Diabetes mellitus. Prevenção. Automonitoramento da Glicemia.

Abstract
Introduction: Systems for self-monitoring of glycemia must provide accurate and reproducible glucose values in order to ensure adequate therapeutic decisions in people with diabetes. Objective: To perform the comparative evaluation between glucose levels, from the determination by remote laboratory test (RLT) and standard laboratory method. Methods: Descriptive study performed with users attended in a Basic Health Unit. Determination of capillary glycemia was performed using Accu-Check Advantage equipment (Roche), using manual lancets, and venous glucose levels were determined with enzymatic methodology (oxidase glucose - Labtest Diagnostic), using quality control of two levels. Results: The determinations of glycemia were performed in 50 volunteers (11 male and 39 female), with a mean age of 61.58 ± 11.08 years, being 25 participants with Diabetes Mellitus type 2. The Spearman correlation coefficient and the Bland Altman test were performed to compare the methodologies. Statistical analyzes showed a strong correlation between levels of capillary and venous glycemia (r = 0.995). Conclusion: The remote laboratory test is effective and may contribute to the control of blood glucose levels, and can also minimize health problems, improving the quality of life of diabetics.

Keywords: Diabetes mellitus. Prevention. Self-monitoring of Glycemia.

Téléchargements

Les données relatives au téléchargement ne sont pas encore disponibles.

Bibliographies de l'auteur-e

Sally Cristina Moutinho Monteiro, Universidade Federal do Maranhão - UFMA

Professora Adjunta de Bioquímica Clínica do Departamento de Farmácia da Universidade Federal do Maranhão - UFMA.

Elizabeth Gomes

Farmacêutico-Bioquímico.

Ilka Kassandra Belfort, Universidade Federal do Maranhão - UFMA

Enfermeira. Mestre em Saúde Materno-Infantil da Universidade Federal do Maranhão - UFMA.

Mauricio Fernandes Avelar

Farmacêutico-Bioquímico.

Romildo Martins Sampaio, Universidade Federal do Maranhão - UFMA

Professor Adjunto do Curso de Engenharia Química da Universidade Federal do Maranhão - UFMA.

Téléchargements

Publié-e

2015-10-26

Comment citer

Monteiro, S. C. M., Gomes, E., Belfort, I. K., Avelar, M. F., & Sampaio, R. M. (2015). ANÁLISE COMPARATIVA DA DETERMINAÇÃO DE GLICEMIA CAPILAR E VENOSA COM GLICOSÍMETRO VERSUS DOSAGEM LABORATORIAL / COMPARATIVE ANALYSIS OF CAPILLARY AND VENOUS GLYCEMIA WITH GLUCOMETER VERSUS LABORATORY DOSAGE. Revista De Pesquisa Em Saúde, 16(1). https://doi.org/10.18764/

Numéro

Rubrique

Artigos Originais / Original Articles